Mindray BS-600M

SKU: BS-600M

Sample continuous loading by racks Up to 120 sample positions long walk-away time. Auto start for sample test. Automatic priority for STAT and rerun.

  • TECHNICAL SPECIFICATIONS

    Throughput 600 photometric tests per hour, up to 800 tests per hour with ISE
    Measuring principles Absorbance Photometry, Turbidimetry
    On-board tests up to 77 photometric tests + 3 ISEs+ 3 serum indices
    Methodology End-point, Fixed-time, Kinetic, ISE

     

    SDM 120 samples by 12 racks, sample continuous loading, STAT in priority
    Sample volume 1.5~45μL, step at 0.1μL
    HbA1c sample Whole blood for HbA1c test
    Sample probe Liquid level detection, clot detection and collision protection

     

    Reagent tray 80 positions with cooling system (2~8°C)
    Reagent volume 10~200μL, step at 0.5μL
    Reagent probe Liquid level detection, bubble detection and collision protection

     

    Sample and Reagent Bar Code Readers
    Support Codabar, ITF (Interleaved Two of Five), Code 128, Code 39, UPC/EAN, and Code 93
    LIS Connectivity Capable of bi-directional communication with LIS (Laboratory Information System)

     

    Reaction volume 70~300μL
    Reaction temperature 37°C with 0.1°C fluctuation
    Reaction cuvettes 124 reusable cuvettes with 8-step auto wash

     

    Light source Tungsten-halogen lamp
    Photometer Grating photometer
    Wavelength 16 wavelengths (340nm, 380nm, 412nm, 450nm, 480nm, 505nm, 546nm, 570nm, 605nm, 630nm, 660nm, 700nm, 740nm, 770nm, 800nm, 850nm)
    Absorbance range : 0~3.5Abs(10mm conversion)

     

    Calibration mode Linear(one-point, two point and multi-point), Logit-Log 4P, Logit-Log 5P, Spline, Exponential, Polynomial, Parabola
    Control rules Westgard multi-rule, Twin plot

     

    Operation system Touch screen(optional), Windows 10(64bit)
    Interface RS-232, Network Port, USB/ parallel port

     

    Power Supply 200~240V, 50/60Hz, 1700VA or 110/115V, 60Hz, 1700VA
    Temperature 15~30°C
    Humidity 35~85%
    Dimension 1380mm(length)*860mm(width)*1200mm(high)
    Weight 350kg

     

  • ADDITIONAL INFORMATION

    Powerful yet Efficient BS-600M

    Flexible Sample Management

    • Sample batch transfer
    • Continuous sample loading by racks
    • STAT Priority
    • Quick rerun

     

    HbA1c Tests with Whole Blood

    • NGSP certified and IFCC traceable
    • Onboard hemolysis without sample preparation
    • No significant interference from Hb variants
    • Lower detection limit of Hb, less rerun rate

     

    Integrated ISE Module

    • Single blood tube for electrolytes and photometric chemistry tests
    • Long-life of ISE electrodes
    • Easy operation and maintenance

     

     

    Reliable Testing System

    • World-class sigma metric performance with original reagent
    • Intelligent abnormal reaction detection, alarming, and auto-rerun to minimize erroneous results

     

    Small Footprint:1.2m2 Photometric 600T/H up to 800T/H with ISE

     

    Flexible Sample Management

    Continuous Sample Loading - Sample continuous loading by racks Up to 120 sample positions long walk-away time

    Quick Start for Sample Rerun - Auto start for sample test.  Automatic priority for STAT and rerun

    STAT Priority - Dedicated STAT position, 1 minute aspiration Quick report ISE+CO2 results in 7 minutes

    Smart Dilution - Patented smart dilution function Faster report

     

    HbA1c Tests with Whole Blood

    Onboard hemolysis with whole blood to improve working efficiency

    No manual pretreatment to reduce biohazardous risks and human errors

    120 T/H for HbA1c test to shorten turn-around-time

     

    Mindray HbA1c reagent with enzymatic method can specifically detect glycated N-terminal amino acids from the hemoglobin beta-chains, which can effectively avoid interference from Hb variants (HbS, HbC, HbE and HbD) and other Hb derivatives (labile HbA1c, carbamylated Hb, and acetylated Hb).

     

    Mindray enzymatic HbA1c gives reliable and cost efficient result with the first measurement and reduce further confirmations.

     

    Lower Hb Detection Limit for HbA1c Test

    Evaluate the Hb concentration in the process of hemolysis and start sample increasement tests in advance for low Hb concentration samples. This function can lower the detection limit and expand linearity range Hb to reduce the rerun rate and reagent waste for anemia samples.

     

    Integrated ISE Module

    • Report Na+, K+, CL- 
    • 1 blood tube for electrolytes and photometric chemistry tests

     

    Efficient and Easy operation

    • Up to 300 tests per hour
    • Up to 9 months of Na/K/Ref electrodes service time
    • Automatic calibration
    • Low consumption
    • Easy operation and maintenance

     

    Reliable Testing System

    Original reagents and validated application protocol lower the inaccuracy and imprecision which contributes to achieving world-class and excellent sigma metric performance.

     

    Higher sigma level system can help laboratories offer high level of quality at reduced cost and enhance clinician confidence and satisfaction.

     

    Intelligent Abnormal Reaction Detection

    Powerful software with in-built algorithms that can monitor the reaction for abnormality, provide flags, and trigger auto-reruns to specimens with abnormal situations in test, such as substrate depletion, oxygen shortage, and hook effects, etc. This intelligent approach with detection, alarming and auto-rerun functions minimizes false results and the risk of a wrong diagnosis, and make reported test result accurate and reliable.