  • TECHNICAL SPECIFICATIONS

    Samples Whole blood or hemolysis sample Number of samples loaded Maximum of 50 samples
    Measurement Parameters HbA1c (Reportable) HbF , HbA1 and eAG (Research) Data storage capacity 100,000
    Measurement principle Ion-exchange-based high-performance liquid chromatography Measurement conditions Ambient temperature: 10℃ - 35℃(Normal operating conditions); -10℃ - 40℃(Storage and transportation conditions) Relative humidity: 20% - 85%(Normal operating conditions); 10% - 90%(Storage and transportation conditions)
    Sample consumption 10uL (Whole blood) 4uL (Pre-diluted mode) Dimensions 650(W)x680(D)x740(H)mm (With auto-loader)
    Accuracy Relative deviation ≤±5.0% Weight ≤77 kg (With auto-loader)
    Reproducibility CV≤1% Column H-10(10,000 tests) 、H-6 (6,000 tests) and H-3 (3,000 tests)
    Linearity range 3.0%~20.1% Filter 3,000test
    Loading mode AL-WB, CT-Predilute, CT-WB Reagent Eluent A, Eluent B and Hemolysis Solution

     

  • ADDITIONAL INFORMATION

    Usability and Parameters

     

    Easy to replace all consumables

    • Automatic sample rotation for barcode scanning
    • Support automated inversion mixing
    • Stand-alone/Track system Support labXpert
    • Closed tube analysis Support cap-piercing of STAT sample

     

    Reagent

    • With RFID tag, automated traceability and convenience
    • Replace the reagent bottles in two simple steps: insert and pull out

     

    Chromatography Column

    • With RFID tag, automated traceability and convenience
    • Replace the column in four simple steps: open, take out, place, and close

     

    High-throughput anti-variants interference

    Test Mode  Throughput ( test per hour)                                  Description
    Standard Mode                         120 Flags 6 hemoglobin variant interferences (HbS/HbC/HbD/HbE/HbJ-Bangkok/HbUbe-2)as "Suspected variant" and automatically switches to variant mode.
    Variant Mode                          60 Resistant to interference from 4 hemoglogin variants (HbS/HbC/HbD/HbE) and provides variant type ags.

     

    Intelligent mode switching for re-testing

    The standard mode and variant mode can be switched automatically, without requiring the replacement of reagents or the column.

     