top of page

Mindray TDR-X 060

SKU: TDR-X 060

TDR-X series automated blood culture systems can dig the data via multi-fields, then generate a excel format file which facilitates customers to do a further analysis.

  • TECHNICAL SPECIFICATIONS

    Sample position 60
    Principle Colorimetry
    Sensitivity ≤10 CFU/ml
    Specimen types Applicable for Blood, CSF, peritoneal fluid, pleural fluid, bone marrow, and other sterile body fluid
    Basic calculation algorithms
    1. Sustained acceleration
    2. Rate
    3. Initial threshold
    Culture type TDR Anaerobic; TDR Resin Aerobic; TDR Resin Peds; TDR Resin Anaerobic
    Extendable Function Not available
    Powerful statistics Export results including bacteria growth curve
    Automation Automatic surveillance and report; Support negative pre diagnostic report
    Temperature controlling system Dual heating systems (solid heating bath and air heating bath)
    Interface USB, RS-232C, Lan
    Storage capacity 500GB
    LIS connection Support BI-LIS
    Dimension and weight Width (600±10mm), Height (525±10 mm), Depth (720±10 mm); 87Kg
    Power supply AC100~240V, 50/60Hz
    Working temperature 0~35℃
    Relative humidity 30~85%, without condensation

     

     

     

     

     

  • ADDITIONAL INFORMATION

    TDR-X series automated blood culture systems can dig the data via multi-fields, then generate a excel format file which facilitates customers to do a further analysis. 

    Department Sending Time Culture Barcode  Culture type Result
    Interal medicine 2014-5-1 11:00 SA14D21001 TDR-Aerobic POS
    ICU 2014-5-5 11:09 SA14D21012 TDR-Aerobic NEG

     

    Drawer type incubator constantly maintains the temperature accurately

    The drawer type incubator can reduce the impacts from external environment, which avoids the false positive readings. Meanwhile, it applies the dual heating system(air bath and solid bath) which keeps the temperature more stable.

     

    One step for loading the culture

    The system can load the culture without touching the screen. It can be finished by scanning and inputting the culture barcode and patient information. The system also supports Bi-LIS mode.

     

    Flexible operation system provides better user experience

    Windows based operation system plus a 12 inches display offers a friendly operation platform

     

    Polythene materials prevent bio-hazard

    The plastic bottle can prevent to be broken by accidental falls, avoiding infection risks.

     

    Variety of resins can reduce the antibiotics interferences

    The resin can not only absorb the antibiotics but also inflammatory factors. This can increases the culture detectable rate. Meanwhile, resin do not interfere the gram stain.

     

    Applicable for many types of specimens

    TDR-X series blood culture systems can be used for testing blood and body fluid(CSF,pleural fluid, etc.)

     

    Various choices of cultures

    TDR-X series can supply standard and resin cultures, both aerobic and anaerobic .

    No. Name No. Name
    1 Resin aerobic culture 1 Pediatric culture
    2 Aerobic culture 2 Resin Anaerobic culture
    3 Resin pediatric culture 3 Anaerobic culture

     

     

    Irreversible colorimetry technology minimizes the problems of delayed specimen

    The blood culture should be sent to laboratory within 2 hours according to CLSI(Clinical and Laboratory Standards Institute) standards. Otherwise, it might affect the bacteria growth. TDR-X series automated blood culture systems can solve the problem by using irreversible colorimetry technology and multi-calculation algorithms.

     

     

     

     

     

Order products like this and more call 305-593-2137 or fill out the form below.