Werfen IL ACL Elite Pro
Routine, specialty and STAT assays on compact, convenient and easy-to-use systems. Offers a unique combination of process automation for the medium-volume laboratory.
ADDITIONAL INFORMATION
Introducing Simple, Comprehensive Solutions for the Hemostasis Laboratory
Routine, specialty and STAT assays on compact, convenient and easy-to-use systems.
ACL ELITE PRO
- Offers a unique combination of process automation for the medium-volume laboratory.
ACL ELITE
- Simplified, offering the same testing capability for the lower volume, less demanding laboratory
ACL ELITE Systems
- Random Access Processing for minimized turnaround time
- STAT PT and APTT available in as little as eight minutes
- Bar-coded reagents (standard on ACL ELITE PRO)
- Superior walk-away capacity for 40 samples and up to 260 on-board cuvettes
- STAT Interrupt and Continuous sample loading for un-interrupted testing capability
- Large integrated color touch-screen
- Intuitive, easy-to-use “Windows®- like” software streamlines system operation
- Minimal mechanical components ensures robustness in design and requires minimal maintenance
- Comprehensive test menu makes specialty testing as simple as routine testing
Flexibility
- Multiple operation modes allow a lab to run the system for maximum efficiency
- 300-test protocol capacity, including 100 customizable by the user
- User-defined reflex testing rules
- User-defined test profiles allow for efficient usage of all reagent positions
- Multiple calibration modes including calibration during
Sample analysis
- Removable tray accommodates sample cups and primary tubes of varying sizes and fill volumes
- On-board bar-code reader ensures accurate sample identification
- STAT samples can be entered at any time in any position on the tray
- One-touch Run starts processing with the press of a button
Reagent Handling
- The ACL ELITE PRO accommodates 22 reagents; the ACL ELITE holds 18
- Reagents can be aspirated directly from the vial
- All reagent positions are monitored for volume and on-board stability
- "Profiling" tests allows a large number of test combinations to be processed within the same analytical session
- Hand-held bar-code reader for reagent vial identification (standard on ACL ELITE PRO)
Wash-R Bottle
One-liter Wash-Reference emulsion; continuous realtime monitoring of liquid level performed through a sensor system.
Dilutors
Precision and accuracy obtained with unique solid-state, fluidic dispensing system.
Sample Tray
Expanded capability holds up to 40 primary tubes/cups and ten additional reagents (from A1 to A10)
Reagent Area
Eight positions dedicated to reagents (from R1 to R8) on ACL ELITE; four additional refrigerated positions available (from R9 to R12) on ACL ELITE PRO.
Disk Drive
Computer-controlled backup, restore, software upgrades, export functions (i.e., patient data, QC data, reaction curves)
CD and Touch-screen
Unbeatable ease of use: just one touch starts the analyzer. Tilts for easy positioning
Rotor Stack
Temperature-controlled and store up to 12 rotors; rotors can be loaded continuously.
Analysis Area
Two channels: 660 nm for clotting assays and 405 nm for chromogenic and immunological assays.
Rotor Movements and Mechanics
Cuvette flow managed by transport and robotic arm mechanisms, allowing walk-away capability.
Probe and Sensors
Two needles for sample and reagent aspiration/ dispensation with capacitance liquid sensors.
Rotor Waste
Container Operator-safe and easily accessible.
Keyboard
In a drawer below the unit.
Adapters Storage
Area For storage of reagent adapters.
Test Analysis
- Centrifugal analysis providing outstanding precision
- Multiple reading systems that optimize analysis efficiency
- Waste area and preheated rotor storage that are continually monitored and allow for emptying and refilling at any time, even if the system is processing samples
- Automated rotor transport from preheat station to analysis area, and then to waste bin when the rotor is completely used, allows increased walk-away capability
Software Capabilities
- Clot signature curves generated through superior optics provide clotting and absorbance tracking for all reactions
- User-defined Rerun and Reflex criteria
- Calibrations can be performed in run along with patient samples
- Customizable system security allows tracking of user actions by lab management
- Disk-drive allows system settings to be exported and saved, along with patient and QC data
- Bi-directional host interface with host query
- Fully automated Factor Parallelism
- Customizable audible alerts
HemosILTM Reagents
- Up to 29 applications are now available on the ACL ELITE and ELITE PRO
- HemosIL reagents are optimized for IL systems
- The exceptional range of sensitivity allows for consistent and superior performances
- HemosIL provides the highest standard of manufacturing and quality control
Routine:
- PT RecombiPlasTin
- PT-Fibrinogen Recombinant
- PT-Fibrinogen-HS Plus
- PT-Fibrinogen-HS
- PT-Fibrinogen
- APTT SynthASil®
- APTT-SP
- APTT SynthAFax®
- APTT Lyophilized Silica*
- Fibrinogen-Clauss
- Thrombin Time (TT)
- Pro-IL-Complex*
- Hepatocomplex*
DVT and PE Diagnosis:
- D-Dimer
Thrombophilia Screening:
- Antithrombin
- Protein-C (chromogenic and clotting)
- Protein S (functional and antigenic)
- Factor V Leiden (APCR-V)
Lupus Anticoagulants:
- LAC Screen and LAC Confirm
- Silica Clotting Time
Coagulation Factors and VWF:
- FII, FV, FVII, FX Deficient Plasma
- FVIII, FIX, FXI, FXII Deficient Plasma
- Chromogenic FVIII
- von Willebrand Factor Antigen
- von Willebrand Factor Activity
Other Specialty Assays:- Heparin
- Plasminogen
- Plasmin Inhibitor
Calibrators and Controls:
- Calibration Plasma
- Normal Control Plasma
- Low Abnormal Control Plasma
- High Abnormal Control Plasma
- Special Test Control Level 1
- Special Test Control Level 2
- D-Dimer Controls (Low and High)
- Low Fibrinogen Control
SYSTEM SPECIFICATIONS
Coagulometric Wavelength 660 nm Chromogenic and Latex Wavelength 405 nm Immunological Tests ✔ Samples On-board (13 x 64 mm, 13 x 75 mm, 13 x 100 mm, 11.5 x 64 mm, 11.5 x 92 mm primary tubes types and cups) Up to 40 Reagents On-board (16, 10, 4 mL vials type) 18 ACL ELITE 22 ACL ELITE PRO Cuvettes On-board 20 ACL ELITE 260 ACL ELITE PRO On-board Bar-code Reader ✔ Sample Predilution ✔ Automated Curve Predilution ✔ STAT Capability ✔ Automated Sample Loading ✔ Automated Rotor Loading N/A ACL ELITE ✔ ACL ELITE PRO Patient Data Storage 30,000 tests - 1,000 samples by 30 tests Quality Control 50 Controls by 20 tests each (750 files by 500 results each) Analytical Reference ✔ Reaction Curves Displayed ✔ Rerun and Reflex Tests 60 rules, 3 criteria each, 10 reflex tests per rule Liquid (sample and reagent) Sensors ✔ Touch-screen (graphic) ✔ Automatic Printing ✔ External Printer Optional (USB or Parallel) ESCP2 Emulation Protocol for the External Printer ✔ HP-PCL Emulation Protocol for the External Printer ✔ ASTM Bi-directional Interface ✔ Automatic Downloading (Host Query) ✔ Automatic Uploading ✔ Mouse ✔ PT/Fib (PT based) Up to 175/hour APTT Up to 125/hour PT/Fib (PT based)/APTT Up to 70 Samples/hour DOWNLOADABLE PDF