Zybio Z50
An evolved 5-diff hematology analyzer
3 histograms for WBC, RBC and PLT
Only 18uL sample volume is required for venous and capillary whole blood test
Only three reagents for CBC + DIFF mode test with optimized liquid handing system
ADDITIONAL INFORMATION
Hematology Analyzer Z50
Z50 is an evolved 5-diff hematology analyzer with compact design that eclipses traditional 5-diff hematology analyzers in the market. It is designed for upgrading your blood analysis solutions, which could provide 29 parameters, 5 scatter-grams, and 3 histograms by advanced semiconductor laser scattering technology, flow cytometry, and impedance method. This Z50 brings you uncompromising performance in medium and satellite laboratories via advanced tri-angle scattering technology, optimal hydraulic system, and streamlined operation procedure.Eclipse traditional 5-diff hematology analyzers
-
Brings you better 5 differentiations of WBC via distinctive tri-angle laser scatter, advanced flow cytometry, and chemical dye, even on high eosinophilic samples.
-
Compared with WBC impedance counting channel, WBC/BASO optical channel can deliver more confident results for basophil measurement.
-
Expanded WBC and PLT linearity range, bringing more accurate results for WBC and PLT, especially for low values.
Enable all possibilities beyond your expectation
• CBC+5 DIFF with 25 reportable parameters and 4 research parameters.
• 3 histograms for WBC, RBC and PLT, 3 2D–scattergrams and 1 3D–scattergram for WBC differentiation, and 1 2D–scattergram for WBC/BASO counting.
• Only 18 µL sample volume is required for venous and capillary whole blood test.
• Only three reagents for CBC+DIFF mode test with optimized liquid handling system, help reduce reagent consumption.
• Only 1 or 2 clicks to access main workflow interface with intuitive software navigation menu and HD touch screen.Enlighten small laboratories
• Innovative reaction system, optimized liquid handling systems, small semi–conductor laser source and two lyses built-in, all lead Z50 to a compact device, fit in small- and middle-labs and clinics.
• Support over 30+ intelligent flags to indicate abnormal blood cell differentiation and morphology.
• One press to shutdown and aspirate probe cleanser for deeply cleaning.Expanded clinical performance
PERFORMANCE
Parameter Background Detection Range CV Carry-over Linearity Range WBC (×10⁹/L) ≤0.2 (3.50~6.90), (7.00~15.00) ≤2.5%,
≤2.0%
≤0.5% (0~500.00) RBC (×10¹²/L) ≤0.02 (3.50~6.80) ≤1.5% ≤0.5% (0~8.00) HGB (g/dL) ≤0.1 (11~18) ≤1.5% ≤0.5% (0~25) HCT (%) / MCV (fL) HCT ≤0.5 MCV (70~120) ≤0.5% / HCT: (0~67) PLT (×10⁹/L) ≤5 (100~149),
(150~500)
≤6.0%, ≤4.0% ≤1.0% (0~5000)
-
SPECIFICATIONS
Principle Flow cytometry + tri-angle laser scatter + chemical dye method for WBC 5-diff differential analysis and WBC counting
Impedance method for RBC and PLT counting
Cyanide free reagent for HGB measurementMode
CBC, CBC+DIF Throughput 60 samples per hour Parameters 25 reportable parameters: WBC, Lym%, Mono%, Neu%, Bas%, Eos%, Lym#, Mono#, Neu#, Bas#, Eos#, RBC,
HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PCT, PDW, PLCC, PLCR
4 research parameters: ALY%, ALY#, LIC%, LIC#Graph 3-D DIFF scatter gram
2-D BASO scatter gram
3 histograms for WBC, RBC and PLTSample Mode
Venous whole blood, Capillary whole blood, Pre-diluted
Sample Volume
Venous/Capillary whole blood: 18 µL
Reagent
Z5 LD Lyse, Z5 LB Lyse, Z5 DN Diluent, Probe cleanser
Calibration Manual calibration
Auto calibration: Fresh blood, calibratorQC Rule L-J, X-B Screen 8.4-inch touch screen Interface 4 USB ports for external printer, external barcode scanner, mouse, keyboard, etc.
1 network port, HL7 protocol, support LISStorage Capacity
≥50,000 results, including numeric and graphical information
Power Supply
(100–240 V)~, 50/60 Hz, 200 VA Operation Temperature: (10–30)°C
Humidity: (20–85)%
Atmospheric Pressure: (70–106) kPaWeight 28.5 kg
Dimension (mm)
230(W) × 455(D) × 435(H) DOWNLOADABLE PDF